ACTA FAC MED NAISS 22009; 26 (1): 11-15 |
Original article
STAPEDOTOMY WITH IMPLANTATION OF THE
FISCH-TYPE 0,4 mm TITANIUM STAPES PROSTHESIS, A GOOD ALTERNATIVE IN STAPES
SURGERY
Stan Cotulbea1,
Alin Hania Marin1, Karina Marin2,
Andreea Steluta Ruja3,
Nikolae Balica1
1ENT Department Timisoara
2Paediatric ENT Department Timisoara
3Odontotherapy and Endodontics Department Timisoara
SUMMARY
The aim of this short-term result study was to present our experience with the
Fisch-type 0,4 mm titanium prosthesis we use in stapes surgery.
We retrospectively analyzed the postoperative hearing results obtained in 162
stapes operations with implantation of the Fisch-type 0.4 mm titanium piston. We
compared them with the hearing results obtained with two types of prostheses we
used before we introduced the titanium piston (49 stapes operation with
implantation of the Causse-type Teflon pistons with a diameter of 0.6 mm and 45
stapes operations with implantation of the Fisch-type Teflon and stainless steel
pistons with a diameter of 0,4 mm). In the majority of patients, a stapedotomy
technique was performed. Only in 14 cases, the stapes footplate was removed.
Pure-tone audiometry with appropriate masking was performed approximately six
weeks postoperatively. The average postoperative ABG was calculated as the
postoperative air PTA minus the postoperative bone PTA.
Although the functional results of the three different types of stapes
prosthesis did not vary significantly, the Fisch-type 0.4 mm titanium piston
gave a somewhat better result in closing the ABG in the “within 10 dB” range.
This may be due to a better fixation of the prosthesis to the long process of
the incus.
The good characteristics of the titanium piston made it our first choice in
stapes surgery, gradually replacing the other two types of stapes prostheses.